CHNS-O-Analysis (CHNS-O)
Analyte
- Organic and inorganic substances in solid and liquid samples
- Sample amount:
- 2 mg normally
- 0.1 – 300 mg (depending on the nature of the sample)
Equipment
- FlashEA 1112 CHNS/O Automatic Elemental Analyzer with 2 Autosamplers MAS200R (Thermo Scientific)
Targets
- Simultaneous quantitative determination of percentage contents of carbon, hydrogen, nitrogen und sulfur
- Quantitative determination of percentage contents oxygen
- measuring range 0.01 – 100 %
- accuracy:
Theoretical value |
Experimental value |
---|---|
100 ppm | 100 ppm ± 10 ppm |
0,10 % | 0,10 % ± 0,01 |
1,00 % | 1,00 % ± 0,02 |
10,00 % | 10,00 % ± 0,1 |
50,00 % | 50,00 % ± 0,3 |
90,00 % | 90,00 % ± 0,3 |
Applications
- Organic elemental analysis
- Protein concentration using N content determination
Measuring principle
- CHNS-Analysis
- O-Analysis
- CHNO-O-Analysis (→ PDF)