Inductively-Coupled Plasma Optical Emission Spectrometry (ICP OES)
Analyte
- Typically liquid aqueous samples
- Solid organic or inorganic samples after micro wave digestion
Equipment
- ICP OES Optima 2100 DV (Perkin-Elmer)
- Mikrowellenaufschlussgerät Start 1500 (MWS Vertriebs GmbH)
Targets
- Multi element analysis - qualitative and quantitative determination of metal ions, Phosphor und Sulfur
Application
- Environmental analysis - water- and effluent water testing
- Trace analysis in core and sludge samples
- Trace analysis in organic und polymer matrices
- Determination of purity
- Quality control
Measuring principle
- Injecting a liquid sample into the nebulizer of the ICP OES
- Introducing the aerosol into an argon plasma
- Formation of atoms and ions at temperatures of 6000 – 8000 K
- Complete atomization of the sample
- Excitation of atoms and ions to emit its characteristic radiation by the argon plasma
- Spectral separation of the emitted light according to wavelength in the spectrometer
- Identifying the elements on the basis of the wavelength of the emitted light
- Quantitative information by measuring the intensity of the emitted light